Specifications

12
NOTE
If the power to furnace is "turned off" or interrupted while in "tuning,"upon returning power to
furnace, the controller display will indicate "LinE FAIL" because sampled data could be
questionable. To restart tuning, refer to "Tuning" procedure.
If the controller cannot maintain temperature setpoint, "tunE FAIL" will appear on display.
First, correct problem responsible for not maintaining the temperature setpoint, then restart
the "tuning" procedure.
Changing Temperature Indication:
Push "SCROLL" button once, "°C" will appear. This indicates temperature measurement. (Contact
Thermolyne if control needs to be changed to °F.)
Setting the High Alarm Temperature (Over Temperature Protection OTP):
Push "SCROLL" button until "AL.SP" (Alarm Setpoint) appears. The lower display should indicate
1100°C.
NOTE
HIAL- HIGH ALARM. (over-temperature protection, OTP)
The controller is fitted with a mechanical relay which is de-energized in the alarm mode. This
relay, when de-energized, removes power from the heating elements. If the primary control
circuit fails, the OTP will control the furnace temperature at the preset value you have
entered. It does not shut off the furnace. but will maintain the chamber temperature at that
value.
Depress either the "UP" or "DOWN" button to select the OTP value you desire. Thermolyne recommends
that you set the value either at the maximum operating temperature of the furnace or a value of 20
degrees above your working temperature if you desire to provide protection for your workload.
Alarm Indications:
Over-temperature
In the event that the furnace chamber temperature exceeds the value of "AL.SP", the red "AL" lamp will
light and the alarm relay will be de-energized, cutting off power to the heating elements. When the
chamber temperature falls below the value of "AL.SP", the red "AL" lamp will extinguish and the alarm
relay will re-energize, returning power to the heating elements and allowing the chamber to heat again.
This process will repeat continuously until you rectify the problem that caused the over-temperature
situation.
Thermocouple Failure
In the event that the thermocouple circuit breaks or the thermocouple fails, the display will indicate
"SnSr FAIL" and the output power level will be decreased to 0%. When the problem with the
thermocouple circuit has been rectified, the controller will resume controlling with the same output power
level used at the moment of the break or failure.