Service manual

126
ROSA messages
5
Service and maintenance
33 Power supply
failure
+15 V error Off Major Disable The supply rail is not within the legal
range. It may compromise the operation.
+3.3 V error Off Major Disable The supply rail is not within the legal
range. It may compromise the operation.
+5.0 V error Off Major Disable The supply rail is not within the legal
range. It may compromise the operation.
-15 V error Off Major Disable The supply rail is not within the legal
range. It may compromise the operation.
51 System
configuration
error
Alarm
queue lost
On Information Disable All messages from previous session(s) are
lost.
Configuratio
nerror
On Information A The detected configuration is different
from what has been defined. Revise your
module configuration.
Default lost On Information C The user-defined default settings are lost.
The decoder starts with factory defaults.
Change the settings, if necessary.
Product
check failed
On Information B A module is found in a place where its
not allowed. Remove the module.
SW options
lost
On Information B The software options codes have to be re-
installed. Use the command SYStem :
OPTion : INSTall.
54 System event Download
started
Off Information Disable The decoder has entered download mode.
Download to a module has started.
16 Video output error Genlock ref.
input loss
Off Warning Disable There is no genlock reference. The video
output runs at nominal frame rate. Check
your genlock reference feed.
Video
decoding
error
On Major A There is no valid video output due to
decoding errors. Check the source and the
condition of the input signal
No. Information Name Enabled Severity Category Explanation/Remedy