User manual
Display settings
80 Nio MKII
Figure 34: DICOM Optimizer dialog
3. Adjust the slider until there is almost no visible difference
between the background of the image above the slider and the
bitmap inside this image.
4. When done, click button Set point 1 of 20.
5. Repeat this procedure until the button is grayed. This indicates the
last point is set.
6. Click OK to finish.
To reset to the latest calibration:
1. Follow step 2 and 3 of the calibration with Barco LCD sensor (see
page 77).
2. In the Calibration dialog (see step no. 4. in the description of the
calibration with sensor), check the option Use the latest
calibration.
3. Click the OK button.
The calibration and the luminance will be reset to the latest
calibrated position.
Note: