User manual

Application settings
95
the system contains at least one display of the same type as
the one being calibrated
at least one of the displays of the same type has already
been calibrated before using a sensor.
To equalize the displays:
1. Be sure the reference display is calibrated to the desired
luminance.
2. Set the Equalization option in the Application Settings dialog.
3. Close the Application Settings dialog.
4. Open the Display Settings dialog and calibrate display (2).
The display will be calibrated and additionally the display
luminance will be matched to the luminance of the reference
display (1).
MediCal Administrator tab
Figure 52: MediCal Administrator tab
Here you can connect to MediCal Administrator, if present.
To connect to MediCal Administrator, click the MediCal
Administrator button. As a result, the MediCal Administrator
connection wizard starts. Please follow the guidelines from the
wizard.
The MediCal Administrator software is a hospital-based softcopy
image quality management system that keeps the consistency
data of every connected display system in a central database. Via