User manual
Display settings
14 NioWatch
2. In the Calibration dialog (see step no. 3. in the description of
the calibration with sensor), check the option Visually and
click the OK button.
The DICOM Optimizer dialog appears.
Figure 9: DICOM Optimizer dialog
3. Adjust the slider until there is almost no visible difference
between the background of the image above the slider and
the bitmap inside this image.
4. When done, click button Set point 1 of 20.
5. Repeat this procedure until the button is grayed. This
indicates the last point is set.
6. Click OK to finish.
To reset to the latest calibration:
1. If you have multiple displays connected (multi-head system),
select the display you wish to reset to the latest calibration.
2. Click the Reset button.