User manual

71
When ready, click on OK.
5 When the calibration is finished successfully, it displays the following
message.
Click on OK to close the window.
Full screen calibration
Full screen calibration is possible on Barco CRT displays only. The optical
sensor must be connected to the display to be able to perform full screen
calibration.
The function is grayed if not available on the display.
The purpose of full screen calibration is to obtain a uniform luminance all
over the screen. When full screen calibration is done, AND the option
Luminance Uniformity Correction (see Luminance Uniformity Correction
on/off Luminance Uniformity Correction on/off) is switched ON, the
luminance will be uniform all over the screen.
If the option Luminance Uniformity Correction is off, the result of the full
screen calibration will not be visible on the screen.
To perform a full screen calibration:
1 In the Configuration section in the main window, right-click on the icon
of the display on which you wish to perform full screen calibration.
From the drop-down menu, select Full Screen Calibration...