Calibrating Barco flat panel displays
Barco flat panel displays, like the Coronis displays, have a built-in optical
sensor that communicates with MediCal Pro.
The calibration runs completely automatically.
When the calibration is finished successfully, it displays the following
message.
Click on OK to close the window.
Calibrating Barco projectors
Barco projectors are calibrated by means of the I-Guard external sensor,
connected to the USB bus. The sensor is included in the Dicom Theater
package.
The calibration runs completely automatically. During calibration, the
same image appears on the flat panel display, but only the projector is
calibrated at that moment.
When the calibration is finished successfully, it displays the following
message.
Click on OK to close the window.
Note: During calibration, be sure no persons or objects come between the
projected sensor patch and the projection screen.