User manual
68
To change the advanced user password:
1 From the main window, press Alt-A on the keyboard or select Tools –
Options – User Login... from the menus. Your user name is still filled
out in the User name field.
2Click on Change Password.
3 In field Old password, fill in the current password.
4 In fields New password and Re-enter new password, fill in the new
password.
5Click on OK. The password is changed.
6 The Login dialog appears again. Click on Cancel to continue as
advanced user, or log in as standard user.
3.4 Calibrating displays
The calibration of the displays is a very important issue. Calibration is
necessary to obtain conformance and to maintain consistency. It is
inevitable during Configuration setup and may be necessary as
intervention when a Q/A check turns out to be Not OK.
You can calibrate Barco displays and non-Barco displays, but the
procedure for both categories is different.
MediCal Pro supports the following sensors for calibration: X-Rite DTP92
(serial or USB), Wellhofer Luxor LXPlus, Solar PMA2200, Minolta LS100 and
Tektronix TJ17. Also the Coronis I-Guard sensor inside BARCO flat panel
displays and the I-Guard sensor from Dicom Theater are supported.
To start calibration:
- Calibration starts automatically after defining a new, uncalibrated
Preset or after modifying a Preset.
- You can also calibrate a display by right-clicking on its icon in the main
window, and selecting Calibrate... from the drop-down menu.