Coronis Fusion 6MP User Guide MDCC-6230 K5902055/02 17/01/2014
Barco nv President Kennedypark 35, 8500 Kortrijk, Belgium Phone: +32 56.23.32.11 Fax: +32 56.26.22.62 Support: www.barco.com/esupport Visit us at the web: www.barco.
Table of contents TABLE OF CONTENTS 1. Welcome! .......................................................................................... 3 1.1 1.2 About the product .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 3 What’s in the box. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 3 2. Parts, controls and connectors .....
Table of contents 7. Repackaging instructions ...................................................................... 35 7.1 Repacking your display.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 36 8. Important information ........................................................................... 37 8.1 8.2 8.3 8.4 8.5 8.6 8.7 8.8 2 Safety information.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..
1. Welcome! 1. WELCOME! 1.1 About the product Overview Thank you for choosing this Coronis Fusion 6MP! Coronis Fusion 6MP is an ingenious PACS display system designed to further enhance flexibility and productivity in diagnostic imaging. Featuring the industry’s first 30-inch color LCD that can be used as two seamless heads or one wide-screen display, Coronis Fusion 6MP offers you the freedom to organize your workspace just the way you want it.
1.
2. Parts, controls and connectors 2. PARTS, CONTROLS AND CONNECTORS 2.
2. Parts, controls and connectors 2.2 Display rear view Overview 1 2 3 4 5 6 7 8 9 10 11 12 Image 2-2 Rear view 1 +24 VDC power input 2 +5 VDC, 0.
3. Display installation 3. DISPLAY INSTALLATION Prior to installing your Coronis Fusion 6MP and connecting all necessary cables, make sure to have a suitable display controller physically installed in your computer. If you are using a Barco display controller, please consult the user guide delivered with it to do this. 3.1 Removing the covers The connector compartment cover should be removed to get access to the connectors. To remove the connector compartment cover 1.
3. Display installation 2. Slide the cover upwards while holding the clips pressed. 3.2 Unlocking the height mechanism In the factory, the height-positioning system in the display stand is locked with a red hook to prevent damage during transportation. You’ll have to remove this hook before adjusting your display height position. To remove the hook: 1. Position the display with its rear side facing you. 2. While holding the display panel pushed down, pull out the red hook in the display stand.
3. Display installation 3. Keep the clip in the dedicated hole in case the display needs to be shipped later. To retain the hook for possible future transportation, insert the short, red end of the hook back into the stand of your display. 3.3 Adjusting the display position Now that the height-positioning system of your display is unlocked, you can adjust the display position. To adjust the display position 1. Tilt, swivel, raise and lower the display as desired.
3. Display installation CAUTION: Do not try to pivot your display when attached to the stand. Trying to do so could cause serious damage to your display and its stand. 3.4 Connecting the signal cables To get access to the connectors, remove the connector compartment cover. See "Removing the covers", page 7 . Each side of your display can have a different video input connected (one side can have DVI connected while the other side has DisplayPort connected).
3. Display installation 3.5 Connecting the power cable To connect the power cable to the display: 1. Connect the supplied external DC power supply to the +24 VDC power input of your Coronis Fusion 6MP display. 2. Plug the other end of the external DC power supply into a grounded power outlet by means of the proper power cord delivered in the packaging. 3.6 Routing the cables To route the cables 1. Route all connected cables through the cable routing channel in the stand of your display. Tip: 3.
3. Display installation Pay attention that all cables stay in the cable channel while re-attaching the cover. To re-attach the connector compartment cover 1. Slide the cover’s top in position and then push the cover’s bottom. You’ll hear a “click” sound of the cover’s clips when the connector compartment cover is in position. 3.8 VESA-mount installation WARNING: Use an arm that is approved by VESA (according to the VESA 100 mm standard). Use an arm that can support the weight of the display.
3. Display installation 1. Gently pull open both lower sides of the cover. 2. Slide the cover upwards while holding the lower sides pulled open. 3. Lift the plastic frame that covers the fixation of the panel to the stand. Turn it for 45 degrees to uncover the fixation screws. 4. Unscrew the four fixation screws fixing the panel to the stand. 5. Attach the panel firmly to the arm using 4 screws M4 x 10 mm.
3. Display installation WARNING: Never move a display attached to an arm by pulling or pushing the display itself. Instead, make sure that the arm is equipped with a VESA approved handle and use this to move the display. Please refer to the instruction manual of the arm for more information and instructions. 3.9 First time starting up Overview You are now ready to start up your Coronis Fusion 6MP for the first time. 1. Switch on your Coronis Fusion 6MP as described in "Stand-by switching", page 16. 2.
4. Daily operation 4. DAILY OPERATION 4.1 Recommendations for daily operation Optimize the lifetime of your display Enabling the Display Power Management System (DPMS) of your display will optimize its diagnostic lifetime by automatically switching off the backlight when the display is not used for a specified period of time. By default, DPMS is enabled on your display, but it also needs to be activated on your workstation. To do this, go to “Power Options Properties” in the “Control Panel”.
4. Daily operation Maximize quality assurance The ’MediCal QAWeb’ system offers online service for high-grade Quality Assurance, providing maximum diagnostic confidence and uptime. Barco recommends to install MediCal QAWeb Agent and apply the default QAWeb policy at least. This policy includes calibration on regular intervals. Connecting to MediCal QAWeb Server offers even more possibilities. Learn more and sign up for the free MediCal QAWeb Essential level at www.barco.com/healthcare/qa. 4.
4. Daily operation Switching on the display while it is in stand-by mode or vice versa can both be done by using the keys or by using the control wheel. To switch your display on using the keys: 1. Light up the keys indicator lights as previously described. 2. While the keys are light up, touch the Stand-by/Cancel key for approximately 2 seconds. As a result, the display will switch on and the power status indicator light is dimmed. To switch your display on using the control wheel: 1.
4. Daily operation How to bring up the OSD menus using the control wheel 1. If not already done so, switch on the display as previously described. 2. Shortly turn the control wheel. As a result, the OSD main menu comes up in the bottom right corner of the screen. If no further actions are taken within the following 90 seconds however, the OSD menu will disappear again. The OSD menu auto-exit function can be disabled in the OSD menu.
5. Advanced operation 5. ADVANCED OPERATION 5.1 OSD menu position About the OSD menu position By default, the OSD menu comes up in the bottom center position of the screen. This position can be modified however. There’s choice from a number of predefined positions. To change the position of the OSD menu: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > User Interface > Menu menu. 3. Enter the Position submenu. 4. Select one of the available OSD menu positions and confirm. 5.
5. Advanced operation 5.4 Power status indicator light About the power status indicator light By default, when the display is switched on, the power status indicator light is dimmed. This behavior can be changed so that the power status indicator light will be blue when the display is switched on. When the display is in stand-by mode, the power status indicator light will always turn orange, even when the power status indicator light is disabled. To enable/disable the power status indicator light: 1.
5. Advanced operation 5.7 Keys sound About keys sound By default, the keys will make a “beep” sound on every touch. However, this sound can be disabled so that no sound will be heard when using the keys. To enable/disable the keys sound: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > User Interface > Controls menu. 3. Enter the Sound submenu. 4. Select Enabled/Disabled as desired and confirm. 5.8 Ethernet over USB About Eth.
5. Advanced operation Enabling the USB endpoint is only recommended when you are using a non-Barco display controller. If you are using a Barco display controller, this communication will automatically be done over the connected video cable(s). To enable/disable USB: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Connectivity menu. 3. Enter the USB submenu. 4. Select Enabled/Disabled as desired and confirm. 5.
5. Advanced operation To enable/disable hibernation on your display: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Power Management menu. 3. Enter the Hibernate submenu. 4. Select Enabled/Disabled as desired and confirm. To specify the hibernate timeout: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Power Management menu. 3. Enter the Hibernate Timeout submenu. 4. Set the timeout value as desired and confirm. 5.
5. Advanced operation 3. Enter the Color Presets submenu. 4. Select one of the available Color Presets and confirm. 5.14 Color temperature About color temperature: It is possible to change the color temperature of your display. Color temperature can only be changed on your display when color presets is set to user. Therefore, please refer to "Color presets", page 23 to do this. To change the color temperature: 1. Bring up the OSD main menu. 2.
5. Advanced operation Diagnostic This mode provides the full calibrated luminance and is intended for using the display for diagnostic purposes. In this mode, the luminance is reduced to approximately half of the luminance. This is intended for using the display with office applications such as word processing. Text Please note that text mode is not persistent, once powered off, the unit will restart in diagnostic mode.
5. Advanced operation 2. Navigate to the Configuration > Calibration menu. 3. Enter the Display Function submenu. 4. Select one of the available display functions and confirm. 5.18 Ambient Light Compensation (ALC) About ALC Ambient Light Compensation (ALC) can only be enabled on your display when the DICOM display function is selected. Therefore, please refer to "Display functions", page 25 to correctly set the display function.
5. Advanced operation Staff Office Corresponds to light conditions in office rooms. Clinical Viewing Room Corresponds to light conditions in diagnostic reading rooms for clinical viewing. Corresponds to light conditions in emergency rooms. Corresponds to light conditions in operating rooms. This setting has the highest maximum ambient light. Emergency Room Operating Room To select a reading room: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Calibration > Ambient Light menu. 3.
5. Advanced operation 5.22 Image source selection modes About image source selection modes Your Coronis Fusion 6MP automatically detects the number of video input signals connected, attaches them to the correct display side and applies the correct video settings to it (resolution, video encoding mode, refresh rate,...). However, it may be needed to manually select the video input signal(s) to be displayed on a certain display side or to adjust certain video settings yourself.
5. Advanced operation To manually select a video input signal: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Image Sources > (Left/Right) Image Source menu. 3. Enter the Input Signal submenu. 4. Select one of the available input signals and confirm. 5.24 Video encoding modes Your Coronis Fusion 6MP display automatically detects the connected video input signals and applies the correct video encoding settings.
5. Advanced operation 5.25 Grayscale conversion modes Your Coronis Fusion 6MP display automatically detects the connected video input signals and applies the correct grayscale conversion settings. Manually selecting a grayscale conversion mode is possible, but then your display’s image source selection mode should be set to allow this. Please refer to "Image source selection modes", page 28 to do this.
5. Advanced operation To manually set EDID timings: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Image Sources > (Left/Right) Image Source menu. 3. Enter the Timings submenu. 4. Select Resolution, Refresh Rate or Color Depth. 5. Select one of the available settings and confirm. 5.27 Display info About display info Your display serial number, color type, native resolution, firmware versions, etc. are available in a dedicated submenu of the OSD menu.
5.
6. Cleaning your display 6. CLEANING YOUR DISPLAY 6.1 Cleaning instructions To clean the display Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized cleaning product for medical equipment. Read and follow all label instructions on the cleaning product. In case of doubt about a certain cleaning product, use plain water.
6.
7. Repackaging instructions 7.
7. Repackaging instructions 3 4 7.1 Bottom buffer Power supply 7 Display controller box 8 Cables (video, USB, power) Repacking your display How to repack your display 1. Place the empty box on a stable surface. 2. Slide the system CD and user guide in the dedicated bag attached at the inside of the box. 3. Place the bottom buffer in the box. 4. Put the power supply in the dedicated cavity of the bottom buffer. 5.
8. Important information 8. IMPORTANT INFORMATION 8.1 Safety information General recommendations Read the safety and operating instructions before operating the device. Retain safety and operating instructions for future reference. Adhere to all warnings on the device and in the operating instructions manual. Follow all instructions for operation and use. Electrical Shock or Fire Hazard To prevent electric shock or fire hazard, do not remove cover. No serviceable parts inside.
8. Important information Power cords: • • • • • Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent, 18 AWG min., rated 250 V min., provided with a hospital-grade type plug 5-15P configuration for 120V application, or 6-15P for 240V application. Do not overload wall outlets and extension cords as this may result in fire or electric shock. Mains lead protection (U.S.
8. Important information human health from uncontrolled waste disposal, please separate these items from other types of waste and recycle them responsibly to promote the sustainable reuse of material resources. For more information about recycling of this product, please contact your local city office or your municipal waste disposal service. For details, please visit the Barco website at: http://www.barco.com/en/AboutBarco/weee Turkey RoHS compliance Türkiye Cumhuriyeti: AEEE Yönetmeliğine Uygundur.
8. Important information 零件项目(名称) 有毒有害物质或元素 Component name Hazardous substances and elements 汞 六价铬 铅 镉 光盘说明书 Pb Hg Cd Cr6+ o o o o 多溴联苯 多溴二苯 醚 PBB PBDE o o CD manual O: 表示该有毒有害物质在该部件所有均质材料中的含量均在 SJ/T 11363-2006 标准规定的限量要求以下. O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement in SJ/T11363-2006. X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出 SJ/T 11363-2006 标准规定的 限量要求.
8. Important information FCC class B This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. This device has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
8. Important information This Coronis Fusion 6MP complies with appropriate medical EMC standards on emissions to, and interference from surrounding equipment. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Interference can be determined by turning the equipment off and on.
8. Important information Immunity test IEC 60601 Compliance level Test levels IEC 61000-4-6 150 kHz to 80 MHz Radiated RF 3 V/m IEC 61000-4-3 80 MHz to 2.5 GHz 3 V/m Electromagnetic environment – guidance equipment should be used no closer to any part of the Coronis Fusion 6MP, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d = 1.2√P d = 1.2√P 80 MHz to 800 MHz d = 2.
8. Important information These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Recommended separation distance The Coronis Fusion 6MP is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
8.
8. Important information Indicates this device must not be thrown in the trash but must be recycled, according to the European WEEE (Waste Electrical and Electronic Equipment) directive Indicates Direct Current (DC) Indicates Alternating Current (AC) Stand-by 8.6 Legal disclaimer Disclaimer notice Although every attempt has been made to achieve technical accuracy in this document, we assume no responsibility for errors that may be found.
8. Important information 8.7 Technical specifications Overview Product acronym MDCC-6230 Screen technology TFT AM Color LCD Dual Domain IPS-Pro Active screen size (diagonal) 772 mm (30.4") Active screen size (H x V) 654 x 409 mm (25.8 x 16.1") 16:10 Aspect ratio (H:V) Resolution native 6MP (3280 x 2048) configurable to 2 x 3MP+ (1640 X 2048) Pixel pitch configurable to 2 x 3MP (1536 X 2048) 0.
8. Important information 20 kg Net weight w/o stand Net weight packaged with 36.
8. Important information POSSIBILITY OF SUCH DAMAGES. MORE INFORMATION/DETAILS IS TO BE FOUND IN EACH SPECIFIC OPEN SOURCE LICENSE. Copyright on each Open Source Software component belongs to the respective initial copyright holder, each additional contributor and/or their respective assignee(s), as may be identified in the respective documentation, source code, README file, or otherwise. You shall not remove or obscure or otherwise alter the respective copyrights.