Specifications

Safety Instructions
15
Power connection - display with external power supply
Power requirements: The display must be powered using the
delivered medical approved 12 VDC power supply.
The medical approved DC power supply must be powered by
the AC mains voltage.
Power cords:
Utilize a UL-listed detachable power cord, 3-wire, type SJ or
equivalent, 18 AWG min., rated 300 V min., provided with a
hospital-grade type plug 5-15P configuration for 120V
application, or 6-15P for 240V application.
Do not overload wall outlets and extension cords as this may
result in fire or electric shock.
Mains lead protection (U.S.: Power cord): Power cords should
be routed so that they are not likely to be walked upon or
pinched by items placed upon or against them, paying
particular attention to cords at plugs and receptacles.
Transient over-voltage
If the equipment is not used for a long time, disconnect it from the AC
inlet to avoid damage by transient over-voltage.
To fully disengage the power to the unit, please disconnect the power
cord from the AC inlet.
Water and moisture
Never expose the display to rain or moisture.
Never use the display near water - e.g. near a bathtub, washbasin,
swimming pool, kitchen sink, laundry tub or in a wet basement.
Ventilation
Do not cover or block the ventilation openings in the cover of the set.
When installing the display in a cupboard or another closed location,
heed the necessary space between the set and the sides of the
cupboard.