User guide
Display information
36
Reading room definition
This function is available only when the DICOM display function is
selected.
To modify a reading room definition:
1. Select the DICOM display function. See To select a display
function (DF):, on page 33.
2. In the ALC & DICOM Options menu, select Reading Room Def.
3. Select the reading room definition you wish to modify. Thereto,
first select Reading Room. Select the appropriate reading room.
4. To modify the maximum ambient light for this reading room,
select Max. Ambient Light and change the value.
You can change this value within certain limits determined by
the selected reading room.
5. To modify the Preset correction value for this reading room,
select Preset Corr. Value and change the value.
You can change this value within certain limits determined by
the selected reading room.
Staff Office Corresponds to light conditions in office
rooms.
Clinical Viewing Corresponds to light conditions in diag-
nostic reading rooms for clinical view-
ing.
Emergency Room Corresponds to light conditions in emer-
gency rooms.
Operating Room Corresponds to light conditions in oper-
ating rooms. This setting has the high-
est maximum ambient light.
Name Description