User guide
Display information
33
• The reading room parameters are pre-defined in the display
according to the AAPM (American Association of Physicists in
Medicine). However they can be changed within certain
limits.
Display function selection
To select a display function (DF):
1. Open the main menu.
2. Open the Display Function menu.
3. Select Display Function.
4. Select the desired DF.
5. Return to the Display Function menu.
You can select the following DFs:
Name Description
DICOM Select a DICOM display function for most
medical viewing applications.
The DICOM function results in more visi-
ble grayscales in the images.
Dynamic Gamma 2.2
Dynamic Gamma 1.8
These are gamma functions that are
shifted to take into account the non-zero
luminance of an LCD panel when driven
with a “black” signal.
They are especially useful in CT applica-
tions to improve the perception of low
Haunsfield values.
Native If you select Native, the native panel
behavior will not be corrected.
Test This is identical to the Native DF.