Coronis 3MP LED Display User Guide MDCG-3221
Coronis 3MP LED Display User Guide MDCG-3221 K5902060/00 10/04/2013
Barco nv President Kennedypark 35, 8500 Kortrijk, Belgium Phone: +32 56.23.32.11 Fax: +32 56.26.22.62 Support: www.barco.com/esupport Visit us at the web: www.barco.
Table of contents TABLE OF CONTENTS 1. Welcome! .......................................................................................... 3 1.1 1.2 1.3 About the product .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 3 Symbols . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 3 What’s in the box. .. .. .. ..
Table of contents 8. Important information ........................................................................... 35 8.1 8.2 8.3 8.4 8.5 8.6 8.7 8.8 2 Safety information.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 35 Environmental information .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 36 Regulatory compliance information .. ..
1. Welcome! 1. WELCOME! 1.1 About the product Overview Thank you for choosing this Coronis 3MP LED Display! Coronis 3MP LED Display is the industry-leading display system for grayscale radiology imaging. It delivers optimum diagnostic precision and workflow efficiency for high-resolution PACS imaging, CT, MRI, chest radiology, angiography, etc. 50 more shades of gray Barco’s high-bright LED backlights render more Just Noticeable Differences to help you see more shades of gray.
1. Welcome! Info, term definition. General info about the term Note: gives extra information about the described subject Tip: gives extra advice about the described subject 1.
2. Parts, controls and connectors 2. PARTS, CONTROLS AND CONNECTORS 2.
2. Parts, controls and connectors 2.
2. Parts, controls and connectors Connector compartment cover open Image 2-3 Rear view with open connector compartment cover 1 +24 VDC power input 5 2 +5 VDC, 0.5A power output 6 Ethernet connector (see note below) USB upstream connector 3 DisplayPort video input 7 USB downstream connector 4 DVI video input The Ethernet connection is used for maintenance purposes and is not supported for user application.
2.
3. Display installation 3. DISPLAY INSTALLATION Prior to installing your Coronis 3MP LED Display and connecting all necessary cables, make sure to have a suitable display controller physically installed in your computer. If you are using a Barco display controller, please consult the user guide delivered with it to do this. 3.1 Unlocking the height mechanism To remove the clip: 1. Position the display with its rear side facing you. 2. Pull the red clip out of the fixation hole in the foot. 3.
3. Display installation 3. To change from portrait to landscape, turn the panel counterclockwise. 4. To change from landscape to portrait, turn clockwise. If, after installing the display of the system, you change the panel orientation while an image is on the screen, the result depends on the graphic board and the resolution of the image. In some cases the image will be rotated automatically, in other cases it will not be rotated (e.g., when pixels would be lost after rotation).
3. Display installation 1. Connect the DVI or DP of the display controller to the DVI or DP connector of the display. 2. If you want to make use of the display’s USB upstream connector, connect a PC USB downstream connector to the display’s USB upstream connector. 3. If you have chosen to use the display’s USB downstream connector, connect a USB device to the downstream connector. 4. Connect the supplied external DC power supply to the +24 Vdc power input of the display. 5.
3. Display installation 3.5 Routing the cables & Reattach the connector compartment cover To route the cables 1. Route all connected cables through the cable routing channel in the stand of your display. Tip: The cable straps at the inside of the connector compartment allow you to fix the cables for better shielding of the cables. To Reattach the connector compartment cover 1. Reattach the connector compartment cover by sliding the cover’s bottom in position and then push the cover’s top.
3. Display installation WARNING: Use an arm that is approved by VESA (according to the VESA 100 mm standard). Use an arm that can support the weight of the display. Refer to the technical specifications of this display for the applicable weight. Overview The panel, standard attached to the tilt & swivel foot, is compatible with the VESA 100 mm standard. So it can be used with an arm stand according to the VESA 100 mm standard. Therefore, the tilt & swivel foot must be removed from the panel. 1.
3. Display installation 5. Remove the four screws fixing the foot while supporting the foot. 6. Attach the arm stand firmly to the panel using 4 screws M4 x 8 mm. 3.7 First time starting up Overview You are now ready to start up your Coronis 3MP LED Display for the first time. 1. Switch on your Coronis 3MP LED Display as described in "Standby switching", page 16. 2. Turn on the computer connected to your display.
4. Daily operation 4. DAILY OPERATION 4.1 Recommendations for daily operation Optimize the lifetime of your display Enabling the Display Power Management System (DPMS) of your display will optimize its diagnostic lifetime by automatically switching off the backlight when the display is not used for a specified period of time. By default, DPMS is enabled on your display, but it also needs to be activated on your workstation. To do this, go to “Power Options Properties” in the “Control Panel”.
4. Daily operation Maximize quality assurance The ’MediCal QAWeb’ system offers online service for high-grade Quality Assurance, providing maximum diagnostic confidence and uptime. Barco recommends to install MediCal QAWeb Agent and apply the default QAWeb policy at least. This policy includes calibration on regular intervals. Connecting to MediCal QAWeb Server offers even more possibilities. Learn more and sign up for the free MediCal QAWeb Essential level at www.barco.com/healthcare/qa. 4.
4. Daily operation 4.4 Bringing up the OSD menus How to bring up the OSD menus The OSD menu allows you to configure different settings to make your Coronis 3MP LED Display fit your needs within your working environment. Also, you can retrieve general information about your display and its current configuration settings through the OSD menu. Bringing up the OSD menus can be done by: 1. If not already done so, switch on the display as previously described. 2. Illuminate the keys as previously described. 3.
4.
5. Advanced operation 5. ADVANCED OPERATION 5.1 OSD menu language About the OSD menu language By default, the OSD menu comes up in English. However, there’s a wide range of other languages available for the OSD menu of your Coronis 3MP LED Display. To change the language of the OSD menu: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > User Interface > Menu menu. 3. Enter the Language submenu. 4. Select one of the available languages and confirm. 5.
5. Advanced operation 5.4 Key indicator lights About the key indicator lights By default, after lighting up, the key indicator lights will dim again if no further actions are taken within the following 5 seconds. However, this behavior can be changed so that the key indicator lights are always on or always off. To configure the key indicator lights 1. Bring up the OSD main menu. 2. Navigate to the Configuration > User Interface > Indicator Lights menu. 3. Enter the Keys submenu. 4.
5. Advanced operation 5.7 USB About USB By default, USB will be enabled. This will allow the connected PC to communicate directly over USB with the internal Coronis 3MP LED Display processor. By disabling USB, communication between the internal Coronis 3MP LED Display processor and the connected PC will not be possible. This does not disable the USB hub and still allows to make use of the display’s USB downstream connectors and any external devices connected to it (keyboard, mouse, ...).
5. Advanced operation 5.9 Hibernate About hibernate Enabling hibernation will not only switch off the backlight but will also force the display to disable other functionalities so that power consumption is further reduced to a minimum. This happens after a specific period of time which can be manually adjusted. Hibernate can only be enabled on your display when the DPMS mode is enabled first. Therefore, please refer to "DPMS mode", page 21 to do this. To enable/disable hibernation on your display: 1.
5. Advanced operation Diagnostic This mode provides the full calibrated luminance and is intended for using the display for diagnostic purposes. In this mode, the luminance is reduced to approximately half of the luminance. This is intended for using the display with office applications such as word processing. Text Please note that text mode is not persistent, once powered off, the unit will restart in diagnostic mode.
5. Advanced operation 2. Navigate to the Configuration > Calibration menu. 3. Enter the Display Function submenu. 4. Select one of the available display functions and confirm. 5.13 Ambient Light Compensation (ALC) About ALC Ambient Light Compensation (ALC) can only be enabled on your display when the DICOM display function is selected. Therefore, please refer to "Display functions", page 23 to correctly set the display function.
5. Advanced operation Staff Office Corresponds to light conditions in office rooms. Clinical Viewing Room Corresponds to light conditions in diagnostic reading rooms for clinical viewing. Corresponds to light conditions in emergency rooms. Corresponds to light conditions in operating rooms. This setting has the highest maximum ambient light. Emergency Room Operating Room To select a reading room: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Calibration > Ambient Light menu. 3.
5. Advanced operation 5.17 Display orientation About Display orientation Your display automatically detects its physical orientation (portrait or landscape) and, by default, automatically adjusts the image orientation to this. This means that when your display is physically rotated, the image will rotate along. The OSD menu however, allows to overrule this behavior and force the image orientation to portrait or landscape regardless of the physical orientation of the display.
5. Advanced operation Standard HDR1 HDR2 Dual 10 (only with Barco display controller) Automatic This mode uses the standard 8 bit encoding mechanism. This mode features 10-bit video encoding following the High Dynamic Range (HDR) standard, usually used on color displays. HDR2 is a Barco specific mode featuring 10-bit video encoding following the High Dynamic Range (HDR) standard but allowing full refresh rate (which is not always possible with HDR1). This mode is usually used on color displays.
5. Advanced operation 5.21 EDID timings About EDID timings Following EDID timings are available for your Coronis 3MP LED Display: Refresh Rate Allows to manually select the refresh rate of the image source video input signal depending on the maximum refresh rate of the display controller connected to your display. Allows to change the orientation of the image source video input signal to landscape, portrait or to let the display automatically assign the correct orientation.
6. Repackaging instructions 6. REPACKAGING INSTRUCTIONS WARNING: Before repacking the display, follow the instruction to replace the protection buffer to prevent damage to the display.
6. Repackaging instructions 3 4 6.1 Protection buffer Display 7 Top box 8 Display controller box Replacing the protection buffer How to replace the protection buffer 1. Place the display on a stable surface. 2. Put the display in the lowest position and fasten the height mechanism, see "Unlocking the height mechanism", page 9 . 3. Very important: Tilt the panel away from the foot before changing the orientation. 4.
6. Repackaging instructions 6.2 Repacking your display How to repack your display 1. Place the empty bottom box on a stable surface. 2. Place the bottom buffer in the box. 3. Put the display together with the protection buffer (see "Replacing the protection buffer", page 30) in the dedicated cavity of the bottom buffer. 4. Put the top buffer on top of the display. 5. Slide the top box over the bottom box. 6. Use the three lock bayonets to lock the box.
6.
7. Cleaning your display 7. CLEANING YOUR DISPLAY 7.1 Front glass Cleaning of the front glass Clean the front glass using a sponge, cleaning cloth or soft tissue, lightly moistened with one of the following tested products: • • • • • • • Flux Windex Glass Plus Bohle glass cleaner Mr. Proper Pril Ajax glass cleaner Sidolin glass cleaner In case none of the above cleaning products is available, use plain water. WARNING: Take care not to damage or scratch the front glass.
7. Cleaning your display The cabinet has been tested for resistance to the following products: - Cidex, Betadine - Alcohol (isopropyl and Ethyl) - Ammonia-based cleaners (Windex) - Aquasonic Gel In case none of the above cleaning products is available, use plain water. Caution: Do not apply or spray liquid directly to the cabinet as excess liquid may cause damage to internal electronics. Instead, apply the liquid to the cleaning cloth. 2. Repeat with water only. 3. Wipe dry with a dry cloth.
8. Important information 8. IMPORTANT INFORMATION 8.1 Safety information General recommendations Read the safety and operating instructions before operating the device. Retain safety and operating instructions for future reference. Adhere to all warnings on the device and in the operating instructions manual. Follow all instructions for operation and use. Electrical Shock • • Do not modify this equipment without authorization of the manufacturer. No user-serviceable part inside.
8. Important information Power cords: • • • Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent, 18 AWG min., rated 250 V min., provided with a hospital-grade type plug 5-15P configuration for 120V application, or 6-15P for 240V application. Do not overload wall outlets and extension cords as this may result in fire or electric shock. Mains lead protection (U.S.
8.
8. Important information 零件项目(名称) 有毒有害物质或元素 Component name Hazardous substances and elements 汞 六价铬 铅 镉 多溴联苯 多溴二苯 醚 Hg Cr6+ Cd Pb PBB PBDE O: 表示该有毒有害物质在该部件所有均质材料中的含量均在 SJ/T 11363-2006 标准规定的限量要求以下. O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement in SJ/T11363-2006. X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出 SJ/T 11363-2006 标准规定的 限量要求.
8. Important information determined by turning the device off and on, the user is encouraged to try to correct the interference by one or more of the following measures: • • • • Reorient or relocate the receiving antenna. Increase the separation between the device and receiver. Connect the device into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help.
8. Important information • • • • Reorient or relocate the receiving antenna or equipment. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced technician for help. Electromagnetic immunity The Coronis 3MP LED Display is intended for use in the electromagnetic environment specified below.
8. Important information Immunity test IEC 60601 Compliance level Test levels Electromagnetic environment – guidance equation applicable to the frequency of the transmitter. Recommended separation distance d = 1.2√P d = 1.2√P 80 MHz to 800 MHz d = 2.3√P 800 MHz to 2.5 Ghz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
8. Important information Recommended separation distance The Coronis 3MP LED Display is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
8.
8. Important information Indicates a medical device that can be broken or damaged if not handled carefully when being stored. Indicates a medical device that needs to be protected from moisture when being stored. Indicates the storage direction of the box. The box must be transported, handled and stored in such a way that the arrows always point upwards. Indicates the maximum number of boxes to be stacked on each other. 15 Indicates that the box should be carried with two persons.
8. Important information Patent information This product is covered under the following intellectual property rights: US Patent RE43,707 US Patent 7,038,186 US Patent 7,166,829 US Patent 6,950,098 European Patent 1 274 066. 8.7 Technical specifications Overview Product acronym MDCG-3221 Screen technology UA-SFT (ultra advanced - super fine technology)+H106 Active screen size (diagonal) 540 mm (21.3”) Active screen size (H x V) 324,9 mm x 433,2 mm (12.79” x 17.
8.
8. Important information 8.8 Open source license information Open source license information Open source license This product contains software components released under an Open Source license. A copy of the source code is available on request by contacting your HC customer support representative at image.careGLOB@barco.com. The software running in the internal processor uses the following licenses: GPL v2.1 and GPL v3.
K5902060-00 April 2013 0120 Barco n.v. President Kennedypark 35 8500 Kortrijk Belgium www.barco.