User guide
8. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5%U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60%dipinU
T
)
for 5 cycles
70% U
T
(30%dipinU
T
)
for 25 cycles
<5%U
T
(>95% dip in
U
T
) for 5s
<5%U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5%U
T
(>95% dip in
U
T
)for5s
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user
of the Coronis 3MP
LED Display requires
continued operation
during power mains
interruptions, it is
recommended that
the Coronis 3MP LED
Display be powered from
an uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3A/m
Not applicable
2
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz
3V
3V/m
Portable and mobile
RF communications
equipment should be
used no closer to any part
of the Coronis 3MP LED
Display, including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d=1.2√P
d=1.2√P 80 MHz to 800
MHz
d=2.3√P 800 MHz to 2.5
Ghz
Where P is the maximum
output power rating
of the transmitter in
watts (W) according
to the transmitter
manufacturer and d
is the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
1. is the a.c. mains voltage prior to application of the test level.
2. Coronis 3MP LED Display doesn’t contain susceptible components to magnetic fields
K5902060 CORONIS 3MP LED DISPLAY 12/01/2015 45