User`s manual
Table Of Contents
- Preliminary Remarks
- Abbreviated summary
- Design and functions
- Operation
- Controlling
- Menu structure
- Maintenance and servicing
- Technical data
- OverView mDG50-DL
- Overview mDR50-DL
- OverView mDR+50-DL
- OverView cDG67-DL
- OverView cDR67-DL
- OverView cDR+67-DL
- OverView cDG80-DL
- OverView cDR+80-DL
- Electrical data illumination unit 120 W
- Optical data illumination unit 120 W, projection unit XGA resolution
- Optical data illumination unit 120 W, projection unit SXGA resolution
- Optical data illumination unit 120 W, projection unit SXGA+ resolution
- Electrical data illumination unit 100 W
- Optical data illumination unit 100 W, projection unit XGA resolution
- Optical data illumination unit 100 W, projection unit SXGA resolution
- Optical data illumination unit 100 W, projection unit SXGA+ resolution
- Optical data illumination unit 200 W, projection unit XGA resolution
- Use of 3200K filter cartridge
- Operating conditions
- Interfaces
- Scope of delivery
- Consumables
- Spare parts
- Controller
- Addresses
- Trouble shooting
- Configuring notes

7. Maintenance and servicing
Barco - OverView D series - Doc-3353-2 - user's manual - Revision 05 - Aug 2005
________________________________________________________________________
7-12
Now you can see the lamp, fixed into its position by means of a securing ring [1].
1
Figure 7-4
inside the lamp housing
• Press the securing ring, and turn it counterclockwise until the hooks are released from the guide rods.
Figure 7-5
the securing ring
• Remove the securing ring
• Withdraw the lamp module. While withdrawing, the plug is disconnected.
Figure 7-6
withdrawing the lamp module
7.3.3.2 Inserting a lamp module
The following description applies if the lamp has already been removed as described above and the lamp cover
is still opened.
• Fetch the lamp module on its housing. The power socket is on the right side.
• Insert the lamp carefully. Take care that the "tubes" slide on the guide rods [2].