User Guide

F-18783-4 © Copyright 2008 TAC All Rights Reserved. 5
CALIBRATION: SINGLE STAGE THERMOSTATS — TC-202
Adjustments
Dial
The dial setting determines the operating point of the thermostat. It may be locked in place
by tightening the recessed Allen set screw in the top of the dial shaft (see Cut 1 in Figure-9).
Differential
The differential between the two contacts of the SPDT switch is easily adjusted by means of
the differential adjustment screw on the thermostat arm (see Cut 2 in Figure-9). This adjust-
ment screw regulates the distance that the thermostat arm is required to move to trip from
one contact to the other. Adjustable from 1° to 15°. Factory set at 2°.
Calibration Procedure
After differential adjustments in the field, the calibration screw (see Cut 3 in Figure-9) may
be used to re-calibrate the control as follows:
1. Immerse the sensing bulb for about 10 minutes in a medium of known temperature.
2. Turn dial to setting corresponding to bulb temperature.
3. Turning calibrating screw to the right (CW) lowers control point temperature, and turning
to the left (CCW) raises control point temperature. Turn the calibrating screw CW or
CCW as required until the switch clicks.
4. Rotate the calibrating screw in the opposite direction until switch clicks again.
5. Set calibrating screw half way between points where switch clicks. (One full turn of
calibrating screw represents approximately 12°.)
Figure-9 Adjustment of Single Stage Thermostats.
Cut 1 Cut 3Cut 2
Figure-10 Part Identification of Single Stage Thermostats.
Dial
Calibrating Screw
Differential
Thermostat Arm
Dial Screw
Bellows Plunger
Cover Screw
Tension Spring
Plunger Spring
Thermal
Switch Insulator
Terminal Block
Switch
Differential
Element
Spring
Adjustment
Screw