Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202522A
Reattaching Hose
1. Attach the cable connector of the new hose to the header plug, making sure that the protrusions on
the connector align with the corresponding slots on the header.
2. Align the at section of the hose end with the corresponding at part of the unit’s hose connector and
slide the replacement hose into the temperature management unit.
3. Screw the hose collar back into place.
4. Reconnect the temperature management unit to a grounded power source.
5. Calibrate operating temperature settings (see page21 for instructions).
6. Perform over-temperature detection system test (see page14 for instructions).