Service manual

English 202430A
25
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202430A
Safety System
 Independent bulb and capillary
  Dual fused input lines
  Over-heat: flashing red light with audible alarm; heater shuts down
 IEC 60601-1; IEC/EN 60601-1-2; UL 60601-1; CAN/CSA-C22.2,
No.601.1; EN 55011; EN 80601-2-35.
 MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO
ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS
ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2,
No.601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No. 60601-1:08;
EN 80601-2-35; Control No.4HZ8
Classified under IEC 60601-1 Guidelines (and other national versions of
the Guidelines) as Class I, Type BF, Ordinary equipment, Continuous
operation. Not suitable for use in the presence of flammable anesthetic
mixtures with air or with oxygen or nitrous oxide. Classified by
Underwriters Laboratories Inc. with respect to electric shock, fire and
mechanical hazards only, in accordance with UL 60601-1, EN 80601-2-35
and in accordance with Canadian/CSA C22.2, No. 601.1. Classified under
the Medical Device Directive as a Class IIb device.
Electrical Characteristics
  Operating speed: 3150 rpm
Airflow: 28-30 cfm
  Peak: 1000W
Average: 450W
  <100µA
  850W Resistive
  15-foot, SJT, 3 cond., 10A
4.6 meter, HAR, 3 cond., 10A
  110-120 VAC, 60 Hz, 9.5 Amperes, or
220-240 VAC, 50 Hz, 4.5 Amperes, or
100 VAC, 50/60 Hz, 10 Amperes
 10A, 200mA (110 - 120 VAC Units)
6.3A, 100mA (220 - 240 VAC Units)
15A, 160mA (100 VAC Units)
 Over-heat test can be performed by the biomedical group.