Service manual
English 202430A
19
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202430A
Normal Operating Temperature Adjustment
WARNING: Improper adjustment of the warming unit’s normal
operating temperature could result in patient exposure to
temperatures outside of the indicated range, and may lead to
patient injury.
• Closelyfollowinstructionsfortemperatureadjustment.
• AlwaysuseaTemperatureTestUnitModel22110toverify
output temperatures.
• AlwaysmakeadjustmentsrelativetotheHIGHtemperature
setting on the warming unit. When the HIGH setting is
accurately adjusted, the MED and LOW settings should also be
within specified limits. If the HIGH range is properly set, and
MED and/or LOW do not meet specs, call Technical Service.
• TheTemperatureTestUnitModel22110doesnotneedtobe
calibrated. The meter it is connected to must be calibrated.
Follow the calibration instructions in the meter service manual
or follow standard facility calibration schedules.
• Whenindoubt,contactTechnicalServiceforassistancein
temperature adjustment.
Tools/Equipment
Temperature Test Unit Model 22110
Small slotted screwdriver
Method
The operating temperature adjustment potentiometer may be
accessed through openings in the upper right corner of the
rear access panel (see Figure 9).
1. Insert the open hose end from the warming unit in the
port on the temperature test device.
NOTE: Do not block the rear air vents on the test device.
2. Turn the warming unit power ON, and allow it to operate
at HIGH for about 3 to 5 minutes before adjusting the
potentiometer.