Service manual

English 202430A
15
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202430A
Over Heat Alarm Thermostat Adjustment
WARNING: Improper adjustment of the warming unit’s over heat
alarm setpoint could result in patient exposure to temperatures
outside of the indicated range, and may lead to patient injury.
• Closelyfollowinstructionsforthermostatadjustment.
• Dangerous voltages are present at the terminals of the
thermostat.
• AlwaysuseaTemperatureTestUnitModel22110toverify
output temperatures.
• TheTemperatureTestUnitModel22110doesnotneedtobe
calibrated. The meter it is connected to must be calibrated.
Follow the calibration instructions in the meter service manual or
follow standard facility calibration schedules.
• Whenindoubt,contact3MPatientWarmingTechnicalSupport
for assistance in temperature adjustment.
Tools/Equipment
• TemperatureTestKitModel22110
• Smallslottedscrewdriver
• Phillips-headscrewdriver
Method
1. Insert the open hose end from the warming unit in the port
on the temperature test device.
NOTE: Do not block the rear air vents on the test device.
2. Connect the warming unit to the power source and turn the
warming unit power ON.
NOTE: The temperature setting will not affect the over heat
adjustment.
3. Using the Phillips-head screwdriver, remove the screw on
the alarm test panel. Lift and remove the panel.