Service manual

Bair Hugger Model 505 Service Manual
8
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202430A
Power Fuse Replacement
The power fuses are located in the power entry module. A
spare fuse is located in the fuse carrier (see Figure 2).
Tools/Equipment
Phillips-head screwdriver
Small slotted screwdriver
Method
1. Disconnect the warming unit from the power source.
2. Remove the 2 Phillips-head screws that secure the cord
retainer. Remove the power cord.
3. Locate the fuse carrier in the center of the power entry
module (see Figure 2).
4. Use the small screwdriver to remove the fuse carrier from
the power entry module.
5. Remove the blown fuse(s) from the fuse carrier.
6. Place the new fuse(s) (as marked) into the fuse carrier.
CAUTION: Replace only with the fuse capacity and type that is
indicated on the adjacent label.
7. Replace the fuse carrier into the power entry module.
8. Reattach the power cord.
9. Replace the 2 Phillips-head screws to reattach the cord
retainer.
10. Reconnect the warming unit to the power source.
11. Record the maintenance action taken and the hours of
operation (read from the hour meter) in the Maintenance
Log section.
Power entry
module (see
detail)
Figure 2. Location
of Power Fuse