ADDITIONAL SCALING PARAMETERS
OTE:N The programming instructions below are only available for loop powered units. Battery powered units will not see
these programming choices.
B2800 Simplied 4-20 mA Programming Instructions
FLOW 4 mA Setting
When the loop powered option is ordered, the flow rate that corresponds to 4 mA must be set. Zero is default flow rate for this
setting. If the current selection is correct, press ENTER once to advance to the next parameter. If adjustment is required, use
the RIGHT arrow to select the position of the number you wish to change. Then use the UP arrow to increment the number.
Once you have completed this step, press ENTER to advance to the next parameter.
FLOW 20 mA Setting
The flow rate that corresponds to 20 mA must be set next. The turbine meter’s maximum flow rate is the default value. If the
current selection is correct, press ENTER once to advance to the next parameter. If adjustment is required, use the RIGHT
arrow to select the position of the number you wish to change. Then use the UP arrow to increment the number. Once you
have completed this step, press ENTER to advance to the next parameter.
4-20 mA Calibration
When ordered with a 4-20mA option, this menu item allows the fine adjustment of the 4-20 mA output. The 4 mA setting
is typically between 35 and 50. To set the 4 mA value, connect an ammeter in series with the loop power supply. At the 4
mA OUT prompt, adjust the 4 mA value to obtain a 4 mA reading on the ammeter. The UP arrow increments the value and
the RIGHT arrow decrements the value. When a steady 4 mA reading is obtained on the ammeter, press ENTER to lock
in this value and move to the 20 mA adjustment. The 20 mA adjustment is performed using the same procedure as the 4
mA adjustment.
4-20 mA Test
The monitor contains a diagnostic routine that allows the simulation of mA values between 4 … 20 to check output tracking.
At the 4-20TEST prompt the arrow keys change the simulated mA output in increments of 1 mA. The ammeter should track
the simulated mA output. If a 4-20 mA test in not necessary, pressing ENTER once will escape the testing at any time.
OTE:N If your B2800 Flow monitor was ordered with a Blancett turbine flow meter, the 4-20 mA was programmed and
factory calibrated.