Figure 4: Typical ammeter connection
4 mA Adjustment
While monitoring the ammeter, adjust the 4 mA value to obtain a 4 mA reading. The UP arrow increments the value and the
RIGHT arrow decrements the value. When a steady 4 mA reading is obtained on the ammeter, press ENTER to lock in this
value and move to the 20 mA adjustment.
20 mA Adjustment
The 20 mA adjustment is performed using the same procedure as the 4 mA adjustment. While monitoring the ammeter,
adjust the 20 mA value to obtain a 20 mA reading. The UP arrow increments the value and the RIGHT arrow decrements
the value. When a steady 20 mA reading is obtained on the ammeter, press ENTER to lock in this value and move to the
next parameter.
4-20 mA Test
The monitor contains a diagnostic routine that allows the simulation of mA values between 4-20 to check output tracking.
At the 4-20TEST prompt the arrows change the simulated mA output increments of 1 mA. The ammeter should track the
simulated mA output. If a 4-20 mA test is not necessary, press ENTER to move to the next parameter.
Linearization
Enhanced accuracy can be obtained by linearization of the display. The linearization routine will accept a maximum of ten
points. Linearization requires additional calibration data from the meter to be used with the monitor. Typically, calibration
information can be obtained in three, five and ten points from the flow meter’s manufacturer. If linearization is not needed,
pressing the RIGHT arrow will take you to the next parameter. (See ADDITIONAL SCALING PARAMETERS on page 16 for
more information.)