TROUBLESHOOTING GUIDE
Trouble Remedy
No LCD Display
Battery Powered Version: Check battery voltage. Should be 1.5V DC. Replace if low or bad.
Loop Powered Version: Check for current flow in the loop. Check polarity of the current loop
connections for proper orientation.
No Rate or Total
Displayed
Check connection from meter pickup to display input terminals.
Check turbine meter rotor for debris. Rotor should spin freely.
Check programming of flow monitor.
Check to see that the minimum flow rate is being met for the current meter in use. Otherwise,
the flow meter will not accurately send pulses to the flow monitor.
Flow Rate Display
Interprets Reading
Constantly
This is usually an indication of external noise. Keep all AC wires separate from DC wires.
Check for large motors close to the meter pickup.
Check for radio antenna in close proximity.
Try disconnecting the pickup from the monitor pig tail. This should stop the noise. If not, then
try reorientating the meter to a new location.
Flow Rate Indicator
Bounces
This usually indicates a weak signal. Replace pickup and/or check all connections.
Examine K factor.
Default K factor Values
Meter Size Default K factor Lower Limit Upper Limit
3/8 in. 20,000 16,000 24,000
1/2 in. 13,000 10,400 15,600
3/4 in. 2750 2200 3300
7⁄8 in. 2686 2148 3223
1 in. 870.0 686.0 1044
1-1/2 in. 330.0 264.0 396.0
2 in. 52.0 41.6 62.0
3 in. 57.0 45.6 68.0
4 in. 29.0 23.2 35.0
6 in. 7.0 5.6 8.0
8 in. 3.0 2.4 4.0
10 in. 1.6 1.3 2.0