User Manual
PROGRAMMING USING PULSE OUTPUT TURBINE FLOW METERS
Each turbine flow meter is shipped with either a K factor value or frequency data. If frequency data is provided, the data must
be converted to a K factor before programming the monitor. K factor information, when supplied, can usually be found on the
neck of the flow meter or stamped on the flow meter body. The K factor represents the number of pulses per unit of volume.
The K factor will be needed to program the monitor readout.
Enter Program Mode
Change to PROGRAM mode by pressing MENU. The mode indicator will change from RUN to PROGRAM.
OTE:N If any input value exceeds the meter’s capabilities for that particular parameter, the LIMIT indicator will begin to flash
indicating an invalid entry. Press ENTER to return to the parameter’s entry screen to reenter the value.
®
Flow Meters
8635 Washington Avenue
Racine, WI 53406
Phone: 800-235-1638 Fax: 262-417-1155
www.Blancett.com
RUN PROGRAM
RELAY1 RELAY2
MENU ENTER
Programming
Keys
Mode
Indicator
Units Display
And
Programming Choices
Rate / Total Display
And
Programming Choices
Figure 3: B2800 panel mount flow monitor
Select Meter Size
At the METER prompt, press ENTER. The current meter size number will begin to flash. Using the arrow keys, scroll through the
size choices to select the bore size of your meter. Press ENTER to save meter size choice.
OTE:N The meter connection size and the bore size are different. For example, many of the one inch NPT turbines have bore
sizes that range from 3⁄8 inch up to 1 inch. Be sure to use the correct bore size or the meter will report incorrect flows
and totals.
Select Display Function
The monitor can display RATE or TOTAL or alternate between BOTH rate and total. At the DISPLAY prompt, press ENTER. The
monitor now shows the display mode currently in effect. To change to an alternate display mode, use the arrows to scroll to
the desired display mode and press ENTER to save the choice.
Select Rate Units of Measure
The monitor allows the choice of many common rate units. (See the SPECIFICATIONS on page 6 for a complete listing of the
unit choices.) At the RATE UNT prompt, press ENTER. The monitor now shows the current rate unit. If the current selection is
correct, press ENTER to advance to the next parameter. To change to an alternate unit, use the arrows to scroll to the desired
rate unit and press ENTER to save the choice.
Flow Monitor, B2800 Panel Mount – Advanced
Page 8 November 2013