POWER SUPPLY One AAA alkaline battery
BATTERY LIFE Approximately 300 tests
WARM UP TIME 10-20 seconds
RESPONSE TIME 10 - 20 seconds
OPERATING
TEMPERATURE
32-104 °F (0-40 °C)
VERIFICATION AND CALIBRATION
The BACtrack is calibrated during manufacture using advanced
alcohol simulation equipment. Known alcohol concentrations
are passed through the sensor to set baseline values for testing.
During testing, the unit compares breath samples to these
baseline values.
The accuracy of breath alcohol testers can fluctuate after six to
twelve months of normal use depending on the number of tests
performed and operating conditions.
If this product is providing inconsistent test results, is not
providing any test results, or provides unusually high or low test
results, the product should be recalibrated immediately by a
trained professional.
The product should be recalibrated at least every six to twelve
months. If the product is used every day, it may need to be
recalibrated as frequently as every month.
15