User manual

sending this as a command in your program. When the subject looks at the
reference position, pressing ‘F9’ key on the Host PC or sending an
“online_dcorr_trigger” command over the link will perform the drift
correction.
Alternatively, an online drift correction can be performed with the aid of a
mouse click. Before recording, add the following line to the FINAL.INI file:
Normal_click_dcorr = ON
Figure 3-20. Performing On-line Drift Correction with
Mouse Click
This will bring up an additional clickable drift correction button in the record
screen. Click on the “Drift Corr” button, which will flash periodically if enabled.
Move the mouse cursor over the intended drift correction target and instruct the
participant to fixate the target precisely. Press the button only once when the
participant fixates stably. The drift correction may fail if there is no stable
fixation data or if there is a large error between the current fixation and the
target item. By default, the maximum acceptable error value (set by the
‘online_dcorr_maxangle’ command) is 5.0°.
3.12 Exiting the Host Application
You can now close the EyeLink 1000 tracker program. Press the key
combination ‘CTRL+ALT+Q’ from any point in the Host PC tracker program to
exit to the command prompt.
An EyeLink 1000 Tutorial: Running an Experiment
© 2005-2008 SR Research Ltd.