User manual
74 An EyeLink 1000 Tutorial: Running an Experiment
© 2005-2008 SR Research Ltd.
To begin the validation procedure, select the “Validate” button or press the ‘V’
key in the Camera Setup screen. The Host PC display will show the gaze
position as a round colored cursor. Note the movements of the cursors, and the
change in relative horizontal position (vergence) following saccades. Once the
cursor appears stable, and close to the target, press the ↵ (ENTER) key to
accept the first fixation. The remaining points are collected automatically or
manually, as in the calibration process.
As each fixation is collected, a cross is used to mark its computed position
relative to the target. The error (in degrees) is printed next to the cross. Similar
to the calibration procedure, the user can use the “Backspace” key in the
middle of a validation sequence to redo data collection for the last or last few
validation points collected. After the final fixation is collected, the average and
worst errors are displayed at the bottom of the screen, and the accuracy is
scored. Each eye is graded separately, using colored messages similar to the
calibration results:
GOOD (green background): Errors are acceptable.
FAIR (grey background): Errors are moderate, calibration should be improved.
POOR: (red background): Errors are too high for useful eye tracking.
Observe the pattern of the errors for each of the targets. If only one target has a
high error, the subject may simply have mis-fixated that point, and the
validation may be repeated to check this: press ‘ESC’ to return to the Camera
Setup screen, and ‘V’ to repeat the validation. If a regular pattern is seen (i.e. all
fixations on the left side are too low) there was probably a calibration or camera
setup problem. In this case, press ‘ESC’ to return to the Camera Setup screen,
and re-calibrate.
3.9 Improving Calibration Quality
The quality of calibrations determines how useful the data recorded will be and
how accurate the gaze calculation will be. Try some of these simple procedures
to improve data quality and gaze accuracy:
• The threshold pupil area must be inside the pupil box (displayed as a red box
around pupil) when the subject is looking at any area of the display. If a
portion of the pupil exits this box, the pupil will be lost.
• The corneal reflection should never be lost or misidentified when the subject
looks around the calibrated area.