User manual

When the last calibration target has been presented, the calibration will be
evaluated. At the bottom of the Calibration screen, each eye's calibration is
graded and displayed as follows:
GOOD (green background): No obvious problems found with the data
FAILED: (red background): Could not use data, calibration must be repeated
The background color of the message indicates the usability of the calibration.
We must still validate the accuracy of the calibration: only serious problems can
be detected here. If problems are found, examine the pattern formed by the
pupil-position cursors (arrays of crosses) for misplaced or missing fixations. If
the calibration was successful, you may press the “Accept” button or the key
to accept the calibration results. Pressing the “Restart” button or the ‘ESC’ key
will restart the calibration. Pressing ‘ESC’ twice exits to the Camera Setup
screen.
Some users (especially the programmers in the phase of testing experiment
programs) may want to run calibration and validation with mouse simulation.
To do this, first delete all of the "M*.cal" files in the “C:\ELCL\EXE” directory of
the Host PC. Start the EyeLink program, set the "Tracking" option as "Mouse
simulation". Go to the Camera Setup screen, type 'C'. This will bring up the
calibration screen. Press the space bar only once to initiate the calibration
process. One cross will be immediately printed on the screen. In addition, the
calibration target and the mouse cursor move to a second calibration point.
Press the left mouse button on the Host PC. Click the left mouse cursor for all
of the following calibration targets, until the calibration finishes.
The Status Panel reports the current status for pupil, corneal-reflection, and
target (EyeLink Remote only) signals and thus will indicate any lapses in
collecting data. In normal operation, the indicators are green. Should any of the
indicators display a color other than green, there is a problem with the setup
that must be addressed to prevent data loss.
Indicates Status of CR-Pupil
OK = Pupil is visible
SIZE = Pupil is too large
MISSING = Pupil is missing
The pupil status error message “SIZE”, highlighted in yellow, indicates that the
size of the pupil is too large or too small. This should not occur when the
camera is mounted on the EyeLink 1000 Tower as the camera to eye distance
has been carefully calculated to ensure compatibility with a wide range of
different pupil sizes. For the EyeLink Remote, the pupil “SIZE” warning typically
suggests that the pupil size is too small because of the ambient lighting or the
eye tracker is placed too far away from the subject.
72 An EyeLink 1000 Tutorial: Running an Experiment
© 2005-2008 SR Research Ltd.