User manual

In general, after threshold adjustment, pupil thresholds should be between 75
and 110 and corneal thresholds should not exceed 230. If the pupil threshold is
too low, try increasing the illumination output. If the pupil threshold or corneal
thresholds are too high, try reducing the illuminator output.
Pupil Clipped and Lost
Good
Corner shadow captures
pupil
Figure 3-17: Corner Effects Seen with Head Rotation
EyeLink 1000 Desktop and Arm Mount Users: If the pupil crosshair flickers
on and off or becomes missing even though the pupil is clearly visible then the
pupil size may be too small, please check the camera distance and the
illumination level. Consider placing the Desktop Mount closer to the subject
(between 40 and 70 cm from the subject’s eye) and/or increasing the
illuminator power level.
3.4 Setting the Corneal Reflection (CR) Threshold
For typical experiments, the “Tracking mode” should always be set to pupil-CR
mode, regardless whether you plan to use the built-in chinrest or not. The
pupil-only mode should only be used together with a bite bar. The corneal
reflection, if present, is identified by a yellow circular shape surrounded by a
crosshair.
Good Corneal Reflection
Poor Corneal Reflection
Figure 3-18: Corneal Reflection
An EyeLink 1000 Tutorial: Running an Experiment
© 2005-2008 SR Research Ltd.
67