4) Click at the blank space next to the graph. A green marker will appear at
intended value. For the ease of adjustments, you may set that point close to
center of the left- and right-eye traces. Also note that a white bar at the
right end of the graph. This bar sets the upper and lower bounds for gain
and offset adjusts.
5) To adjust the gain of eye traces, place the mouse cursor outside of the
regions bounds by the white bar. Drag the mouse up or down until the span
of the upper and lower eye traces spans 8 volts. You will notice that both
the gain and offset values are updated when you drag the mouse up or
down.
6) Now, place the mouse cursor in the regions bounds by the white bar. Drag
the mouse up or down until the top of the eye trace is aligned with 4 volts
and bottom of the eye trace is aligned with -4 volts. Repeat steps 5) and 6)
for fine tuning.
7) Once you are happy with the adjustments, don’t forget to unselect the “SEL”
buttons on the “Adjust” button so that you will not accidentally touch the
adjustments.
8) Now your “Calibration” is done. Click on the “Pause” button to continue
recording.