User manual

2.3.4 Calibrate Screen
28 EyeLink 1000 Host Application
© 2005-2008 SR Research Ltd.
2.3.4.1 Calibrate Screen Purpose
Calibration is used to collect fixation samples from known target points in order
to map raw eye data to either gaze position or HREF data. Targets are serially
presented by the Display PC. The participant fixates each while samples are
collected and feedback graphics are presented on the Host PC display. The
calibration is automatically checked when finished, and diagnostics are
provided. Calibration should be performed after camera setup and before
Validation. Validation provides the experimenter with information about
calibration accuracy.
The zoomed and global views of the camera image, along with pupil and CR
threshold values, are displayed at the bottom left of the screen. The eye to be
calibrated as well as the calibration type (as defined in the Set Options screen
or via the EyeLink API) is indicated beside the camera images at the bottom of
the screen. The calibration status and current calibration target being
presented are indicated on the bottom right of the screen.
To perform a calibration, have the participant look at the first fixation point and
select the ’Accept Fixation‘ button, ENTER or the Spacebar, to start the
calibration. If ‘Auto Trigger’ is disabled (‘Force Manual Accept’ from the Set
Options screen is enabled), repeat this action after each target fixation.
Performing one of these actions quickly, twice in succession, can switch from an
automatic calibration to forcing a manual calibration at any point into the
calibration sequence. This can be useful for subjects showing difficulty fixating
targets or who inappropriately anticipate new target positions.
Figure 2-5 Calibrate Screen