26 EyeLink 1000 Host Application
© 2005-2008 SR Research Ltd.
Click Help (F1) to access the online help page for
Camera Setup. All available key shortcuts are listed
on the Help screen.
Keyboard Shortcuts: F1 = open Help screen
Click ‘Calibrate’ to go to the Calibrate screen. After
setting up the camera and adjusting thresholds
(EyeLink 1000 and EyeLink 2000) or biases (EyeLink Remote), you need to
calibrate the system.
Keyboard Shortcuts: C = go to Calibrate screen
Click ‘Validate’ to go to the Validate screen. Validation
shows the experimenter the gaze position accuracy
achieved by the current calibration parameters. Validation should always be
run after Calibration.
Keyboard Shortcuts: V = go to Validate screen
or Click ‘Drift Correct’ to go to
the Drift Correct screen. A Drift Correction/Check is recommended before each
trial to ensure accuracy of the calibration parameters is maintained. Generally
this is initiated via the application running on the display PC.
Keyboard Shortcuts: D = go to Drift Correct screen
Click to go to the Video Setup screen. See “EyeLink
Video Overlay Option User’s Manual” for details. This
button is useful only if your system has been licensed for video overlay option.
Keyboard Shortcuts: W = Video overlay configuration.