User manual

Camera Setup Screen Main Functions 2.3.3.2
Click “Auto Threshold” to have the Host PC compute
threshold levels automatically. Fine tuning may be
necessary.
The EyeLink Remote dynamically adjusts threshold levels that are further
biased by threshold adjustments.
Keyboard Shortcuts: A = Auto Threshold selected image
Clicking these buttons manually increases or decreases the
selected pupil threshold (Tower or Desktop Mounts) or pupil
threshold biases (EyeLink Remote).
Keyboard Shortcuts: and = increase and decrease pupil
threshold/bias
In Pupil-CR mode, these buttons manually increase or decrease the
selected CR threshold (Tower or Desktop Mounts) or CR bias
(EyeLink Remote).
Keyboard Shortcuts: + and - = increase and decrease CR threshold/bias
Select the tracking mode for recording. Typically,
with most shipped systems, Pupil-CR is the only
mode available as Pupil alone tracking requires complete head immobilization
for high accuracy.
Keyboard Shortcuts: P = toggle Pupil only or Pupil-CR mode where possible
Select the sampling rate for recording. Here 1000
Hz is selected. The 2000 Hz rate is available only
with the EyeLink 2000 system.
Keyboard Shortcuts: F = alternates Sample Rate selection
Selects the method used to fit the pupil center and
determine pupil position. Measure of Area or
Diameter are always based on the Centroid Model.
An Ellipse model is the only method available with the EyeLink Remote option.
Keyboard Shortcuts: Q = toggle selected pupil shape model
Toggles display of pupil and CR crosshairs in camera
images.
Keyboard Shortcuts: X = toggle crosshair display on/off
EyeLink 1000 Host Application
© 2005-2008 SR Research Ltd.
23