User manual
Select whether to record eye events
(fixations and saccades) in Gaze or
HREF coordinate. GAZE is screen gaze
x, y; HREF is head referenced-calibrated x, y
Keyboard Shortcuts: E = alternates between Gaze and HREF settings
Defines the sensitivity of the
EyeLink 1000 parser for saccade
event generation. Normal is intended for cognitive tasks like reading; while High
is intended for psychophysical tasks where small saccades must be detected.
See Section 4.3.5 Saccadic Thresholds for details of event parsing.
Keyboard Shortcuts: X = alternates between Saccade Sensitivity levels
Select filter level of data
recorded to the EDF file.
Each increase in filter level reduces noise by a factor of 2 to 3.
Keyboard Shortcuts: F2 = alternates between filter levels for the EDF file
Note: Online Parsing and the EyeLink Data Viewer assume use of the File
Sample Filter. SR Research Ltd recommends leaving this value set to EXTRA.
Select filter level for data
available via the Ethernet
link.
Each increase in filter level reduces noise by a factor of 2 to 3 but introduces a
1-sample delay to the link sample feed.
Keyboard Shortcuts: F3 = alternates between filter levels for the link
Select type of data for analog
output. OFF turns off analog
output; PUPIL is raw pupil x, y;
HREF is head referenced-
calibrated x, y; GAZE is screen gaze x, y
Keyboard Shortcuts: A = alternates between analog output options
18 EyeLink 1000 Host Application
© 2005-2008 SR Research Ltd.