User manual
If enabled, requires the manual pressing of the
spacebar or ENTER key on Host or Display PC
in order to gather the sample when the
subject is looking at each calibration or validation target. If disabled, the
calibration and validation procedure automatically samples once the eye settles
on a target.
Keyboard Shortcuts: Y = toggle Force Manual Accept on/off.
Clicking the Camera Position Detect button
polls the position of the camera selection knob
to determine the eye to track. This option is only available for the Tower Mount.
Keyboard Shortcuts: K = toggle camera eye autodetect on or off.
Lock the setting of the eye to record eye on the
Display PC keyboard if performing a monocular
recording. This option is applicable to the binocular Desktop Mount only.
Keyboard Shortcuts: K = Lock the currently selected Eye.
Search Limits, though not recommended for most subjects, are useful for
images with pupil or CR foils, such as reflections off of glasses or makeup.
Search Limits delimit the area of the camera image (indicated by a red
boundary in the Host PC global view) to be examined for the pupil or CR in the
event that tracking of these is lost. A red shape around the searched area
appears in the Host PC’s global view.
If “Search Limits” is selected
and the pupil position is
moved, search
for pupil is confined to the
area within this box; otherwise, the whole image is searched for pupil.
If Move Limits is checked, the search limit box moves with the pupil.
Search Limits are automatically active with the EyeLink Remote as the entire
image is searched on every frame given that head movement is not restricted.
In Mouse Simulation mode the Host PC mouse simulates eye movement and
can be used for experiment testing and debugging purposes.
Keyboard Shortcuts: M = toggle on/off Mouse Simulation, F4 = toggle Search
Limit on/off, F5 = toggle dynamic updating of the Search Limit area around the
pupil
Record the participants’ eye area or
diameter in pixels. The area is
recorded in scaled image pixels. Diameter is calculated from pupil area fit using
a circle model.
Keyboard Shortcuts: S = alternates between pupil Area or Diameter data