User manual

Important Information
© 2005-2008 SR Research Ltd.
115
6.2.3.1
6.2.3.2 Cables:
Re-mount the new illuminator <instructions depend on the mount>, restore
the routing of the illuminator cable(s), and reconnect the cable to the camera or
illuminator.
Reconnect the power supply, start the application software, and check to be
sure the illuminator is producing proper output. If the mount allows the angle
of the illuminator to be adjusted, it may be necessary to adjust the angle of the
illuminator to provide the best illumination of the object of interest.
6.2.3 Cables and Lenses:
The following components are replaceable, if the substitutions are made
with components supplied by or approved by SR Research Ltd.
Camera Lenses:
Almost any C-mount lens may be used in the visible spectrum. The high-
resolution sensor in the EyeLink CL camera performs best with high-resolution
machine-vision lenses, and noticeable blurring may be seen at the edges of the
image with standard CCTV optics. Please contact SR Research if a CS-mount
lens is required for your application.
The EyeLink CL camera is optimized for near-infrared (to 910 nm) use,
however performance in this range depends critically on proper lens selection.
The majority of C-mount lenses perform poorly beyond 800 nm, with blurry
images (especially zoom lenses) or dark images due to loss from optical
coatings. Please contact SR Research for a current list of lenses we have
approved for IR use.
The illuminator cables should not be replaced with other cables without the
express permission of SR Research. The ferrites on the cable must remain in
place and be positioned within 10 cm (4”) of the camera.
The Camera cable (from the computer to the camera) may be replaced with
any compatible cable, up to 10 meters in length.