User manual
112 Important Information
© 2005-2008 SR Research Ltd.
6. Important Information
6.1 Safety
6.1.1 Eye Illumination Safety
WARNING: Illuminators must only be connected to EyeLink CL camera, and
only the supplied cables may be used.
CAUTION: Use of a controls or adjustments or performance of procedures other
than those specified herein may result in hazardous radiation exposure.
CLASS 1 LED DEVICE
IEC 60825-1 (Ed. 1.2:2001)
The EyeLink CL illuminators are compliant with the IEC-60825-1 LED
safety standard as a Class 1 LED device. This standard has been or is in the
process of being adopted by most countries, and regulates many aspects of LED
and laser eye safety, including retinal, corneal and skin safety. Class 1 products
are “safe under reasonably foreseeable conditions of operation, including the
use of optical instruments for intrabeam viewing”.
As these illuminators may be used in situations where they may be viewed
for long periods of time, some precautions should be observed. Our experience
has been that even safe levels of IR illumination can eventually cause some
discomfort due to the slight drying effect of even this low level of illumination.
(This is especially true for wearers of contact lenses). SR Research recommends
that the illuminators not be used for extended periods at distances of less than
150mm (6 inches) from the eyes. This will ensure an exposure of less than 1.2
mW/cm² (milliwatts per square centimeter). Exposure decreases as the square
of the distance, so even slightly larger distances will reduce exposure
significantly.