Specifications

Mi180 measures 6 different parameters: pH, ORP, EC, TDS
(Total Dissolved Solids), percentage of NaCl and tempera-
ture in a variety of ranges.
pH calibration can be performed in 3 points selectable
between 7 memorized buffers, to provide a very accurate
calibration curve even when testing different samples, where
very wide differences in pH can be found.
The auto-ranging feature for EC and
TDS measurements automatically sets
the resolution suitable to the tested
sample. All measurements can be tem-
perature compensated at 20 or 25°C
and the compensation coefficient is
selectable by the user.
The automatic temperature compensa-
tion can also be disabled for measuring
the actual conductivity value. The stability
indicator on the LCD ensures accuracy.
Conductivity readings are
performed with the 4-ring
probe supplied with the
meter.
The GLP feature allows users to
store and recall data on system
status.
PC compatible through an RS232
port or USB.
Ordering Information
Mi180 is supplied complete with
MA917B/1 pH Electrode
MA814DB/1 EC/TDS/NaCl/Temperature probe
MA831R Temperature Probe
MA9315 Electrode Holder
M10004 pH 4.01 Sachet Buffer solution
M10007 pH 7.01 Sachet Buffer solution
M10010 pH 10.01 Sachet Buffer solution
M10030 12880 µS/cm calibration solution
M10031 1413 µS/cm calibration solution
M10016 Sachet Electrode Cleaning solution
Mi5200 Application Software
MA9350 RS232 connection cable
with 2 meters cable
Graduate Pipet, 12 VDC Adapter & Instruction
manual
Accessories
MA9004 pH 4.01 buffer, 230 mL bottle
MA9007 pH 7.01 buffer, 230 mL bottle
MA9010 pH 10.01 buffer, 230 mL bottle
MA9015 Electrode storage solution, 230 mL bottle
MA9016 Electrode cleaning solution, 230 mL bottle
MA9112 pH 12.45 buffer solution, 230 mL bottle
MA9060 12880 µS/cm calibration solution,
230 mL bottle
MA9061 1413 µS/cm calibration solution,
230 mL bottle
MA9063 84 µS/cm calibration solution,
230 mL bottle
MA9065 111.8 mS/cm calibration solution,
230 mL bottle
MA9066 100% NaCl calibration solution, 230 mL bottle
MA9069 5000 µS/cm solution, 230 mL bottle
MA9310 12 VDC Adapter, 220 V
MA9311 12 VDC Adapter, 110 V
MA9315 Electrode Holder
MA917B/1 Double junction refillable pH electrode
MA814DB/1 EC/TDS/NaCl/Temperature probe
with DIN connector and 1 m cable
MA921B/1 Platinum ORP electrode with 1 m
cable (will be replaced by SE310)
SE310(*) Platinum ORP electrode with 1 m
cable
MA831R Temperature probe
MA9350 RS232 connection cable with
2 meters cable
(*) Available from the 1
st
of September 2010
Specifications Mi180
Range pH -2.00 to 16.00 pH; -2.000 to 16.000 pH
mV ±699.9 mV; ±2000 mV
EC 0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm;
3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm (uncompensed EC*)
TDS 0.0 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm);
1.50 to 14.99 g/L (ppt); 15.0 to 100.0 g/L (ppt);
up to 400.0 g/L actual TDS (with 0.80 factor)
NaCl 0.0 to 400.0%
Temp -20.0 to 120.0°C / -4.0 to 248.0°F
Resolution pH 0.01 pH; 0.001 pH
mV 0.1 mV; 1 mV
EC 0.01 µS/cm; 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm;
TDS 0.01 mg/L; 0.1 mg/L; 1.0 mg/L; 0.01 g/L; 0.1 g/L
NaCl 0.1%
Temp 0.1°C / 0.1°F
Accuracy pH ±0.01 pH; ±0.002 pH
mV ±0.2 mV; ±1 mV
EC ±1% of reading ±(0.05 µS/cm or 1 digit)
TDS ±1% of reading ±(0.03 ppm or 1 digit)
NaCl ±1% reading
Temp ±0.4°C / ±0.8°F
Rel mV offset ±2000 mV
Calibration pH 1, 2 or 3 points calibration, with 7 memorized buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18,
10.01 and 12.45)
EC 1 point slope calibration with 6 memorized solutions: (84 µS/cm, 1413 µS/cm,
5.00 mS/cm, 12.88 µS/cm, 80.0 µS/cm, 111.8 mS/cm)
NaCl 1 point, with MA9066 solution
Temp 2 point, at 0 and 50°C / 32 and 122°F
Temperature Compensation automatic or manual, from -20.0 to 120.0°C / -4.0 to 248.0°F
Temperature Coefficient selectable from 0.00 to 6.00%/°C (EC and TDS only)
pH Electrodes & Temp Probe MA917B/1 & MA831R (included)
EC/TDS/NaCl/Temp Probe MA814DB/1 (included)
TDS Factor 0.40 to 0.80 (default value is 0.50)
Log on demand up to 50 samples on each range (pH, mV, EC, TDS, NaCl)
GLP last pH, EC, NaCl calibration data
PC Interface RS232 / USB Opto-isolated
Environment 0 to 50°C / 32 and 122°F; max RH 95%
Input Impedance 10
12
Ohm
Power supply 12 VDC power adapter (included)
Dimensions 230 x 160 x 95 mm
Weight 0.9 kg
Mi180
pH/ORP/EC/TDS/NaCl/Temperature Laboratory Bench Meter
MULTI
32
www.milwaukeeinst.com
(*) Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation.