Specifications
These simple 1-point calibration testers are designed for the
Horticultural and Agricultural market.
They are built in a one-piece splash-proof casing and they
are immersible up to the LCD.
All models have Automatic Temperature Compensation and
the temperature sensor housing is made out of 316 inox
steel instead of plastic.
This ensures 100 times faster readings and higher accuracy.
Models available are:
• C61: Conductivity tester low range
Range: 0 to 1999 µS/cm
• C62: Conductivity tester high range
Range: 0.00 to 19.99 mS/cm
• T71: TDS tester low range
Range: 0 to 1999 ppm
• T72: TDS tester high range
Range: 0.00 to 10.00 g/L (ppt)
C61/C62/T71/T72
High accuracy Conductivity and TDS Testers
Specifications
C61 C62 T71 T72
Sharp EC Sharp EC Sharp TDS Sharp TDS
Range 0 to 1999 µS/cm 0.00 to 19.99 mS/cm 0 to 1999 ppm 0.00 to 10.00 g/L
Resolution 1 µS/cm 0.01 mS/cm 1 ppm 0.01 g/L
Accuracy ±2% Full Scale ±2% Full Scale ±2% Full Scale ±2% Full Scale
Temperature Compensation automatic from 5 to 50°C automatic from 5 to 50°C automatic from 5 to 50°C automatic from 5 to 50°C
Battery Life 350 hours 350 hours 350 hours 350 hours
Dimensions 155 x 45 x 25 mm 155 x 45 x 25 mm 155 x 45 x 25 mm 155 x 45 x 25 mm
Stainless steel casing ATC
Temperature has a huge influence on TDS and EC, on the basis of a 2% error per degree
Celsius. Even small temperature differences between the measured solution and the
measuring instrument can lead to large reading errors. For instance, if the solution's tem-
perature is 15°C, and the tester’s is 35°C, the reading error is (35-15x2%) 40%.
Conventional testers with plastic casings for their temperature sensors also slow response
times. It can take at least 10 minutes for conventional testers to fully compensate
temperature differences. This causes reading error.
Sharp tester's new stainless steel encased temperature sensors can compensate in
seconds, no matter how large the temperature difference. Sharp tester’s fast response
time guarantees high accuracy.
316 stainless steel casing
Plastic casing
Conventional tester
Sharp Tester
Ordering Information
C61 and C62 are supplied complete with protective cap,
20 mL 1413 µS/cm sachet of calibration solution, hard
carrying case, calibration screwdriver, batteries and
instructions.
T71 and T72 are supplied complete with protective cap,
20 mL 1382 ppm sachet of calibration solution, hard car-
rying case, calibration screwdriver, batteries and instruc-
tions.
Accessories
M10030B 12880 µS/cm calibration
solution, 20 mL (25 pcs)
M10031B 1413 µS/cm calibration
solution, 20 mL (25 pcs)
M10032B 1382 ppm (mg/L) calibration
solution, 20 mL (25 pcs)
M10038B 6.44 ppt (g/L) calibration solution,
20 mL (25 pcs)
M10080B 800 ppm calibration solution
20 mL (25 pcs)
Temperature sensor
EC/TDS
26
www.milwaukeeinst.com