Specifications

EC/TDS
20
www.milwaukeeinst.com
Mi170 measures 4 different parameters - EC, TDS (Total
Dissolved Solids), percentage of NaCl and temperature in
a variety of ranges.
The auto-ranging feature for EC and TDS measurements
automatically sets the resolution suitable to the
tested sample. All measurements
can be temperature compensated
at 20 or 25°C and the compensa-
tion coefficient is selectable by
the user.
The automatic temperature com-
pensation can also be disabled
for measuring the actual conduc-
tivity value. The stability indicator
on the LCD ensures accuracy.
Conductivity readings are performed
with the 4-ring probe supplied with the
meter. The GLP feature allows users to
store and recall data on system status.
PC compatible through an RS232 or USB port.
Mi170
Autoranging EC/TDS/NaCl/Temperature Laboratory Bench Meter
More accurate readings with the
4-RING MA814DB/1 EC/TDS/NaCl
and Temperature probe!
Conductivity readings are performed
by applying an alternate current
to the 4-ring probe which
creates a variable volt-
age depending on
the conductivity.
Ordering Information
Mi170 is supplied complete with
MA814DB/1 EC/TDS/NaCl/Temperature Probe
MA9315 Electrode Holder
M10030 12880 µS/cm calibration solution
M10031 1413 µS/cm calibration solution
Mi5200 Application Software
MA9350 RS232 connection cable with 2 meters cable
• 12 VDC Adapter
• Instruction manual
Specifications Mi170
Range EC 0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm;
30.0 to 200.0 mS/cm; up to 500.0 mS/cm actual conductivity (uncompensated EC)*
TDS 0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm);
1.5 to 14.99 g/L (ppt); 15.0 to 100.0 g/L (ppt); up to 400.0 g/L actual TDS* (with 0.80 factor)
NaCl 0.0 to 400.0%
Temp -20.0 to 120.0°C / -4.0 to 248.0°F
Resolution EC 0.01 µS/cm; 0.1 µS/cm; 1.0 µS/cm; 0.01 mS/cm; 0.1 mS/cm
TDS 0.01 mg/L; 0.1 mg/L; 1.0 mg/L; 0.01 g/L; 0.1 g/L
NaCl 0.1%
Temp 0.1°C / 0.1°F
Accuracy EC ±1% of reading ±(0.05 µS/cm or 1 digit)
TDS ±1% of reading ±(0.03 mg/L or 1 digit)
NaCl ±1% of reading
Temp ±0.4°C / ±0.8°F
Calibration EC 1 point slope calibration with 6 memorized solutions (84.0 µS/cm, 1413 µS/cm,
5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm)
NaCl 1 point, with MA9066 calibration solution
Temp 2 points, 0 to 50°C / 32 to 12 °F
Temp. Compensation automatic or manual, from -20.0 to 120.0°C / -4.0 to 248.0°F
Temp. Coefficient selectable from 0.00 to 6.00%/°C (EC and TDS only)
Probe MA814DB/1 4-ring probe with built-in temperature sensor (included)
TDS Factor 0.40 to 0.80 (default value is 0.50)
Log on Demand up to 50 samples on each range (EC, TDS, NaCl)
GLP last EC, NaCl calibration data
PC Interface RS232 / USB Opto-isolated
Environment 0 to 50°C / 32 to 122°F; max RH 95%
Power supply 12 VDC power adapter (included)
Dimensions 230 x 160 x 95 mm
Weight 0.9 kg
Accessories
MA814DB/1 EC/Temperature probe with DIN
connector and 1 m cable
MA9060 12880 µS/cm calibration solution,
230 mL bottle
MA9061 1413 µS/cm calibration solution,
230 mL bottle
MA9063 84 µS/cm calibration solution,
230 mL bottle
MA9064 80000 µS/cm conductivity solution,
230 mL bottle
MA9065 111.8 mS/cm calibration solution,
230 mL bottle
MA9066 100% NaCl calibration solution,
230 mL bottle
MA9069 5000 µS/cm solution, 230 mL bottle
MA9310 12 VDC Adapter, 220 V
MA9311 12 VDC Adapter, 110 V
MA9315 Electrode holder
MA9350 RS232 connection cable with
2 meters cable
Mi5200 Application Software
Rear Connector Panel layout
Communication to the PC is done via opto-isolated USB
and RS232 ports.
(*) Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation.