User`s guide
Physios
Physio Troubleshooting
Revision D.0 5-15
ssn February 10, 1999 C:\WINNT\Profiles\dapowell\Desktop\D.0 Books\CD FILES SONOS
2D reference image
does not update
Make sure 2D Hold or BMode Hold is off in CW and
PW modes.
Verify that is off. Verify that the ECG leads are
connected properly.
Adjust Delay or Beats rotary control if an R-wave is
present, or Interval rotary control if there is no R-wave
(all Physio Trigger controls).
Adjust ECG Gain rotary control.
System does not
stop beeping
Touch RWave Beep to turn off the beep during each
R-wave.
Waveform shape is
wrong
Make sure each physio cable is secure in the correct
connection on the system, and that touch controls
match your selections.
For example, for ECG and Pulse waveforms, Physio
Setup should show ECG and Pulse set to Normal to
obtain the signal from a physio transducer connected to
the system, or to AUX for an auxiliary signal from
another source.
Symptoms Suggestions
Freeze