User`s guide

Physios
Physio Troubleshooting
5-14 Revision D.0
Physio Troubleshooting
Always make sure that the active preset is appropriate for the study being
performed.
Symptoms Suggestions
Physio does not
display.
For physios other than ECG, make sure Physio
Suppress is off.
Check to ensure the physio cable is secure in the
correct connection and is correctly attached to the
patient.
Touch Test to test physio signals.
For ECG and respiration signals, check the patient
cable, the connection to the system, the electrodes, and
the lead placement.
Adjust Gain and Position controls for the active
waveform. If there is no touch control for the physio
waveform you want to view, press , and make
sure this physio is set to Normal (for a physio
transducer connected to the system) or Aux (for an
external signal connected).
Too many physios
display.
With Physio active, press , and set the physios
you do not want to see to Off.
Waveform is too big
or too small.
Adjust the physio Gain and Position controls for the
active waveform.
The ECG or
respiratory
waveform contains
an artifact or
wanders
Follow the electrode manufacturers suggestions for
skin preparation.
Put a little gel on the electrodes. They might be dry.
Reconnect the lead to the electrode, and check the
electrode placement on the patient.