User`s guide

Physios
Controls
Revision D.0 5-5
ssn February 10, 1999 C:\WINNT\Profiles\dapowell\Desktop\D.0 Books\CD FILES SONOS
Trigger Selects the triggering mode: Off, ECG, Timer, Loop/
ECG or Loop/Timr. Trigger activates either Beats and
Delay or time Interval for triggering 2D, Color, Angio,
or Doppler (with or without spectral) updates.
Loop/ECG and Loop/Timr are available in Contrast
or AD only.
With Contrast active, Loop/ECG lets you acquire the
same portion of the cardiac cycle while the 2D, Color,
or Angio image is displayed in real time. With AD
active, Loop/ECG lets you acquire images while the
2D image is displayed in real time. Updates the
acquisition based on the selected beats and delay.
With Contrast active, Loop/Timr lets you acquire
images at a slower interval while the 2D, Color, or
Angio image is displayed in real time. With AD active,
Loop/Timr lets you acquire images while the 2D
image is displayed in real time. Updates acquisition in
milliseconds, based on selected intervals.
When PW is active, turning Trigger off activates
simultaneous Duplex Doppler mode.