User`s guide

The System
Setting Up the System
1-14 Revision D.0
NOTE While not a direct danger to patient or operator safety, ambient magnetic fields
may affect the colors that are displayed on the imaging screen. To remove the
effects of this interference, periodically degauss the monitor. Push up on the
degaussing button located underneath the front-right corner of the monitor case.
Calibrating monitors with brightness/contrast push buttons on the front
2. On the monitor, turn the brightness dial down just until the bottom
grayscale bar (the 16th bar) looks pure black and disappears into the
screen background. When the monitor is properly calibrated, 15
grayscale bars are visible, running from white (top) to nearly black
(bottom).
6apc0158
1. On the monitor, press and hold the right contrast button only until the
top grayscale bar looks pure, bright white. Once it looks pure white,
release the button.
Tip: Adjust monitor
lighting when the
ambient room lighting
changes, such as for
bedside studies.
Black grayscale bars
White grayscale bar
+
++
++
++
+
-
--
--
--
-