User`s guide
Imaging Modalities
Angio Imaging
7-112 Revision D.0
Secondary
Adaptive Angio Changes the Angio frequency as a function of focal
depth. When the Angio box is at the top (shallow
depth), the highest frequency is displayed, and vice-
versa.
Angio
Priority
Displays all received Angio data by giving higher
priority to Angio and overriding the BMode
information.
Backgnd Fills the entire color box with the lowest Angio flow
velocity color.
Colorize Optimizes contrast resolution by activating the current
colorization map and overlaying the grayscale image.
To change the active map, turn the color rotary control
just under the Colorize touch control.
LGCs (Lateral Gain Controls) Each of the eight sliders adjusts
the amplification of returning signals within a specific
lateral 2D/BMode or AQ image area. Available only
with sector and curvilinear array transducers.
LGC Touch control that turns the Lateral Gain Controls on
and off.
Magnify Increases or decreases the magnification factor of an
image in Zoom or Zoom Preview modes. Appears as a
secondary control when Contrast is on.
Map Changes the Angio flow map used for real-time and
looping. Provides a selection for user preferences.
Packet Selects Angio packet size: S (small), M (medium), or L
(large).
Persist Adjusts the amount of frame averaging, which can give
a smoother appearance.
Power Adjusts transmit power in decibels (dB) where
0.0dB = maximum power and -30.0dB = minimum
power. Is a primary right-touch-panel control when