User`s guide
Imaging Modalities
PW and CW Imaging
7-106 Revision D.0
Signal is audible, 
but cannot see it.
Make sure the monitor is adjusted for the ambient 
light. See “Calibrating the Monitor” on page 1-13 for 
details.
Gain to display more Doppler information.
Compress and  Reject to increase the range of 
echoes displayed.
Filter.
For cardiac, readjust the transducer position to get 
more parallel to flow. For vascular, use Cursor Angle 
to adjust the estimated angle of flow.
Turn on Colorize.
Press   and make sure Baseline Suppress is off.
Make sure that ADP is turned on.
Spectrum is noisy. Gain to display less Doppler information.
Filter to increase the hertz level below which 
audible signals are removed (fewer signals display).
Reject to increase the amount of low-level signal 
removed from the spectrum.
Compress to eliminate weaker signals and noise 
from the spectrum.
Make sure the sample volume is as small as possible.
Press . Temporal Smooth and  Velocity 
Smooth. Also make sure Baseline Suppress is on.
Area around the 
baseline is 
blackened out.
Filter to decrease the hertz level below which 
audible signals are removed (more signals display).
Press   and make sure Baseline Suppress is off.
Spectrum is 
speckled.
Press . Temporal Smooth and  Velocity 
Smooth.
Symptoms Suggestions
Setup
Setup
Setup
Setup










