User`s guide
Imaging Modalities
PW and CW Imaging
Revision D.0 7-89
ssn February 10, 1999 C:\WINNT\Profiles\dapowell\Desktop\D.0 Books\CD FILES SONOS 
D.0\System Basics D.0\Frame Files\7CH.FM add.2
Sweep Changes the sweep speed for the Doppler spectrum. Is 
displayed only when Spectral is on. Available speeds 
are 25, 50, 100, and 150 mm/sec.
Threshold Determines the type of signals used in calculating the 
trace waveform. A higher setting indicates lower level 
pixels are used to calculate the trace waveform. To use, 
first press 
 .
Trace [Trace Auto, Trace Top, Trace Bottom] 
Works in conjunction with autotrace’s blue line. Press 
the   key and then touch Trace Auto. You can 
then toggle through the trace selections. Trace Auto 
includes PW information which is seen above and 
below the baseline. Trace Top displays the signal 
above the baseline, and Trace Bottom displays the 
signal below the baseline.
Triplex Activates simultaneous triplex Doppler mode, in which 
both the image and Doppler spectral trace are live with 
Color/Angio on. Available in vascular presets.
Waveform
Suppress
Suppresses waveform line used for autotrace after 
.
Zoom
Turns BMode zoom (magnification) on or off. To move 
the zoom preview area, use the trackball (  is 
active when you first enter zoom preview). To resize 
the zoom preview area, press   and use the 
trackball. Then touch Zoom again to enter full zoom 
mode.
Freeze
Freeze
Freeze
Tip: When activated, 
Zoom is first outlined 
by a box. This 
indicates zoom 
preview. Touch Zoom 
a second time to go 
into full zoom. 
Position
Size