Imaging Modalities
Color Imaging
7-66 Revision D.0
Noncardiac primary
With Color active, touch PW, CW, or MMode to go into PW Color, CW Color,
or MMode Color (combination) mode. The second mode control (PW, CW, or
MMode) is highlighted, and the Color touch control is outlined by a box,
indicating that Color mode is still active. To exit the combination mode, touch
Color. Now the Color touch control is highlighted, indicating that only Color
mode is active.
Power Adjusts transmit power in decibels (dB) where
0.0dB = maximum power and -30.0dB = minimum
power. Is displayed as a primary right-touch-panel
control set when Contrast is on.
Secondary
Controls
Switches between the primary and secondary touch
panels. When the control is highlighted, the secondary
controls are active.
Smoothing Adjusts the amount of color scan line averaging,
affecting the smoothness and sensitivity of the color
display.
Baseline Unwraps aliased signals. Shows higher velocities
flowing in one direction by lowering color assignments
for velocities flowing in the other direction.
B/W Suppress Suppresses the black-and-white image that appears
outside of the color image, thereby increasing the frame
rate. Available with sector and curvilinear array
transducers.
Color Compare Displays grayscale and color versions of an image side
by side.
Color Suppress Removes color from an image while remaining in
Color mode. Frame rate is unaffected.
Density Trades off resolution versus frame rate by varying line
density.