User`s guide
Imaging Modalities
2D/BMode Imaging
7-32 Revision D.0
Controls
Primary
BMode is used for all presets except cardiac.
Colorize Optimizes contrast resolution by activating the current 
colorization map and overlaying the grayscale image. 
To change the active map, turn the color rotary control 
just under the Colorize touch control.
Focal Strength When on, the focus is increased at the selected focal 
zone or zones, producing better resolution in the area of 
interest. If more that one focal zone is selected, the 
focal carets are automatically positioned closer together 
on the screen, indicating the increased focal intensity. 
(Focal Strength is only available on linear and CLA 
transducers.)
Focal Zones Selects up to four possible transmit focal zones. The 
actual number of zones available is dependent on 
transducer and depth. A caret appears on the screen for 
each selected focal zone. As focal zones are increased, 
the frame rate decreases. 
Focus Repositions the acoustic depth of the 2D/BMode focal 
zone, indicated by carets.
Frequency Fusion Optimizes frequencies for penetration, texture, or 
resolution. Changes are reflected in the Frequency 
Fusion icon.
Harmonic Fusion 
(Touch Control)
Activates Harmonic Fusion for imaging tissue. 
Activates Harmonics for a 2D image only. Available 
only with a harmonics-ready transducer, such as the s3, 
s4, s8, 11-3L, c3540, OmniPlane III, or x4.
Harmonic Fusion 
(Rotary Control)
Selects the Harmonic Fusion setting when Harmonic 
Fusion is on. Changes are reflected in the Harmonic 
Fusion icon.